Analytical method development and validation for Estimation of Calcium Dobesilate and Docusate Sodium in Pharmaceutical dosage form with QbD approach
Jaydeep S Chauhan, Ritika Gajre
Sigma Institute of Pharmacy Bakrol, Vadodara 390019.
*Corresponding Author E-mail: jaydeepchauhan090499@gmail.com
ABSTRACT:
Analytical method development and validation for estimation of Calcium Dobesilate and Docusate Sodium in Pharmaceutical dosage form with QbD approach. Combination of Calcium Dobesilate and Docusate Sodium helps for the treatment of Haemorrhoid. There is no any UV spectroscopy method of this combination due to less absorbance of the Docusate Sodium in various solvent. A specific, precise and accurate RP-HPLC method has been developed and validated for estimation of Calcium Dobesilate and Docusate Sodium in Capsule dosage form. Apply QbD approach by using statistical tool Design of experiment with Box-Behnken design for the optimization of the screening method of RP-HPLC method. Calcium Dobesilate and Docusate Sodium were estimated on Thermoscientificsyncronis C18 (250mm X 4.6mm), 5μm column using Acetonitrile: 0.01M Tetrabutylammonium dihydrogen phosphate pH-4.8(66:34, v/v) as mobile phase with flow rate 1.0 ml/min and detection was carried out at 214 nm. The retention time of Calcium Dobesilate and Docusate Sodium were found to be 2.943 min and 4.190 min respectively. The linearity and range was found to be 50-450 μg/mL for Calcium Dobesilate and 10-90μg/mL for Docusate Sodium. % RSD of precision was found to be less than 2%. %RSD for Robustness parameters was found to be less than 2%. % Recovery of Calcium Dobesilate and Docusate Sodium at different levels were found in the range of 99.451% - 100.082% and 98.071% - 101.187% respectively. So, the developed method was precise, robust and accurate. The assay value for Calcium Dobesilate and Docusate Sodium was found to be 99.772% - 99.874% and 98.159% - 98.736% respectively.
KEYWORDS: Calcium Dobesilate, Docusate Sodium, RP-HPLC method and QbD with Design of Experiment.
1. INRODUCTION
ANALYTICAL METHOD DEVELOPMENT [1,2]
Pharmaceutical products formulated with more than one drug, typically referred to as combination products, are intended to meet previously unmet patients need by combining the therapeutic effects of two or more drugs in one product. These combination products can present challenges to the analytical chemist responsible for the development and validation of analytical methods. The development and validation of analytical methods (Spectrophotometric), High performance liquid chromatography (HPLC) and High performance thin layer chromatography (HPTLC) for drug products containing more than one active ingredient. The official test methods that result from these processes are used by quality control laboratories to ensure the identity, purity, potency of drug products.
NEED FOR DRUG ANALYSIS
The reasons for the development of newer methods of drug analysis are:
· No official method is available in IP, BP and USP for estimation of drug or combination of drugs.
· A proper analytical procedure for the drug may not be available in the literature due to patent regulations.
· Analytical methods for a drug in combination with other may not be available.
· Analytical methods may be available for the drug in the form of a formulation due to the interference caused by the formulation excipient.
· Analytical methods for the quantitation of the drug in biological fluids may not be available.
QUALITY BY DESIGN [3,4]
The pharmaceutical Quality by design is a systemic approach to development that begins with predefined objective and emphasizes product and process understanding and process control, based on sound science and quality risk management. Quality by Design is emerging to enhance the assurance of safe, effective drug supply to the consumer, and also offer promise to significantly improve manufacturing quality performance.
Experimental design is a powerful technique and tool for QbD, used for exploring new processes, gaining increased knowledge of the existing processes and optimizing these processes for achieving internationally competitive performance.
DRUG PROFILE
CALCIUM DOBESILATE
TABLE 1: DRUG PROFILE: CALCIUM DOBESILATE: [5,12]
Name |
CALCIUM DOBESILATE |
|
IUPAC Name |
Calcium 2,5 dihydroxybenzenesulfonate |
|
Description |
White and solid state. It is the calcium salt of dobesilic acid. |
|
Structure |
|
|
Categories |
Vasoprotective agent |
|
CAS Number: |
20123-80-2 |
|
Molecular weight |
436.4gm/mole |
|
Chemical Formula |
C12H10CaO10S2 |
|
Properties:- |
||
State |
Solid |
|
Melting point |
>300°C |
|
Solubility
|
Solvent |
Solubility |
Water |
Soluble |
|
Ethanol |
Soluble |
|
2-propanol |
Poorly soluble |
|
pKa |
7.67 |
|
Log P |
1.20 ± 0.71 |
|
Dosage form |
Capsule |
|
Storage |
Store in a well closed container. Protect from Sunlight. |
|
Mechanism of Action |
Calcium dobesilate acts on the endothelial layer and basement membrane of the capillaries. It reduces histamine and bradykinin-induced hyper permeability. It increases red blood cell membrane flexibility and reduces capillary fragility. Calcium dobesilate can reduce the platelet aggregation stimulated by collagen and thrombin. |
|
Pharmacokinetic Absorption
|
Absorption: It is mainly absorbed from the G.I.T. and achieves a peak serum concentration (P.S.C) of 8 mcg/ml in 6 hours. Distribution: Calcium Dobesilate exhibits 25% Protein Binding and is distributed throughout the body except the brain since it does not seem to cross the BBB (Blood Brain Barrier). Metabolism: It is metabolised and excreted by the kidneys. Excretion: 75% unchanged in Urine, 25% in faces in 24 hours; Calcium Dobesilate achieves plasma half Life in 4.9 hours. |
|
Metabolism |
It is metabolised and excreted by the kidneys. |
DOCUSATE SODIUM
TABLE 2: DRUG PROFILE: DOCUSATE SODIUM: [7,8,13]
Name |
Docusate sodium |
|
IUPAC Name |
2- Sodium 1,4-bis(2-ethylhexoxy)-1,4-dioxobutane-2-sulfonate |
|
Description |
White to off-white, and hydroscopic. It is also known as dioctyl sodium sulfosuccinate. |
|
Structure
|
|
|
Categories |
Purgative or laxative |
|
CAS Number: |
577-11-7 |
|
Molecular weight |
444.6 gm/mol |
|
Chemical Formula |
C20H37NaO7S |
|
Properties:- |
|
|
State |
Solid |
|
Melting point |
153-157°C |
|
Solubility
|
Solvent |
Value |
Acetone |
Soluble |
|
Ethanol |
Soluble |
|
Water |
Sparingly soluble |
|
pKa |
6.0 |
|
Log P |
5.96 ± 0.52 |
|
Storage |
Docusate Sodium should be stored in tightly closed Containers at temperatures below 24°C. |
|
Dosage form |
Capsule |
|
Mechanism of Action |
Reduces surface tension of the oil-water interface of the stool resulting in enhanced incorporation of water and fat allowing for stool softening. |
|
Pharmaco-kinetic Absorption |
Onset of action :12 to 72 h (oral); 2 to 15 min (enema). |
2 LITERATURE REVIEW:
OFFICIAL METHODS FOR CALCIUM DOBESILATE:
TABLE 3: OFFICIAL METHODS FOR CALCIUM DOBESILATE
Sr.No.
|
Drug |
Method |
Description |
Ref. No. |
(1) |
Calcium Dobesilate monohydrate (I.P. /B.P.) |
Potentiometric titration |
Titrate: 0.2gm calcium dobesilate+10ml water+40ml dil.H2SO4 Titrate: 0.1M cerium sulphate |
5,12 |
REPORTED METHODS FOR CALCIUM DOBESILATE:
1. Detection of Calcium Dobesilate in calcium Dobesilate capsules: a comparison of HPLC and UV spectrophotometry. [13]
2. Determination of Calcium Dobesilate in human plasma using ion-pairing extraction and high-performance liquid chromatography. [14]
3. Method development and validation for the simultaneous quantitative estimation of Calcium Dobesilate and Troxerutin in tablets by reverse phase HPLC. [15]
4. Method development and validation for the simultaneous quantitative estimation of Calcium Dobesilate and Troxerutin in tablets by reverse phase HPLC. [16]
OFFICIAL METHODS FOR DOCUSATE SODIUM:
TABLE 4: OFFICIAL METHODS FOR DOCUSATE SODIUM
Sr.No. |
Drug |
Method |
Description |
Ref. No. |
(1) |
Docusate Sodium (I.P. /B.P.) |
Acid –Base titration |
Titrant: 1.0mg Docusate Sodium+25ml 0.5M alcoholic potassium hydroxide (reflux 45min) Titrate: 0.5M HCl Indicator:0.25ml phenolphthalein solution |
7,12 |
(2) |
Docusate Sodium (U.S.P.) |
Iodimetric titration |
Titrant: 50mg Docusate Sodium+50ml chloroform+50ml salt solution Titrant: Tetra-n-butyl ammonium iodide Indicator: 0.5ml bromophenol blue TS |
8 |
(3) |
Docusate Sodium (CAPSULE) (U.S.P)
|
RP-HPLC |
Column: C18(15 cm X 4.6mm; 5μ) column Mobile Phase: Acetonitrile: 0.01 M Tetrabutylammonium hydrogen phosphate (66: 34, v/v) Detection wavelength(UV): 214 nm |
8 |
(4) |
Docusate Sodium (TABLET)(I.P/U.S.P) |
RP-HPLC |
Column: C18(10 cm X 4.6mm; 5μ) column Mobile Phase: Acetonitrile: 7mM ammonium acetate, (50:50, v/v) Detection wavelength(UV): 210nm |
7 8 |
AIM AND OBJECTIVE AND RATIONALE:
AIM: Analytical method development and validation for simultaneous estimation of Calcium Dobesilate and Docusate Sodium in capsule along with QbD approach.
OBJECTIVE:
· To develop UV spectroscopic method for the simultaneous estimation of Calcium Dobesilate and Docusate Sodium.
· To develop precise, accurate and selective HPLC Method for the simultaneous estimation of Calcium Dobesilate and Docusate Sodium.
· To validate developed HPLC method as per ICH guideline.
RATIONALE:
· Literature Reviews suggest that there are several analytical methods for Calcium Dobesilate and Docusate sodium alone.
· Literature Reviews also suggest that there are several analytical methods for other combination with Calcium Dobesilate and Docusate sodium.
· But no HPLC and UV method are reported for simultaneous estimation of Calcium Dobesilate and Docusate sodium in their combined Dosage form.
So, The Rationale of my work is to develop selective, precise and accurate analytical method for the estimation of Calcium Dobesilate and Docusate sodium in their combined dosage form.
EXPERIMENTAL WORK
MATERIALS AND METHODS
Instruments and apparatus
TABLE 5: INSTRUMENT USED
SR. NO. |
NAME OF INSTRUMENTS
|
MAKE |
MODEL |
1 |
Analytical balance |
Mettler Toledo, Switzerland |
ML204/A01 |
2 |
U.V. Visible Spectrophotometer |
Shimadzu, Japan |
UV 1800 UV 1700 |
3 |
HPLC – Gradient |
Shimadzu, Japan |
Detector: SPD-20A Pump: LC-20AT |
RP-HPLC HPLC – Isocratic |
Detector: SPD-20A Pump: LC-20AD |
||
4 |
pH Meter |
Digital instrument corporate |
LI 120 |
5 |
FTIR Spectrophotometer |
Bruker, Germany |
ALPHA |
6 |
Ultra sonicator |
Fronline FS 4 Mumbai, India |
-- |
7 |
Triple Distilled Water |
Borosil |
-- |
8 |
Melting Point |
Veego |
VMP-D |
Apparatus: Volumetric flasks, pipettes, beakers, measuring cylinder are of durasil and borosil. Mobile phase bottles are of Schott (duran), Germany, and Calibrated Thermometer.
Materials and reagents
Materials: -Calcium Dobesilate, Docusate Sodium
Reagents: All the chemicals and reagents were used of only AR and HPLC grade. Details are as shown in Table 6
TABLE 6: CHEMICALS AND REAGENTS
SR NO. |
NAME OF CHEMICALS/ REAGENTS |
MANUFACTURER |
1 |
Calcium Dobesilate |
Bajaj pharmaceutical Industry, Vadodara, Gujarat |
2 |
Docusate Sodium |
Bajaj pharmaceutical Industry, Vadodara, Gujarat |
3 |
Methanol (HPLC Grade) |
Fisher Scientific, Mumbai |
4 |
Acetonitrile (HPLC Grade) |
Fisher Scientific, Mumbai |
5 |
Ortho-phosphoric Acid |
Fisher Scientific, Mumbai |
6 |
Tri-ehtylamine |
Fisher Scientific, Mumbai |
7 |
Tetrabutyl ammonium dihydrogen phosphate |
SDfine |
IDENTIFICATION OF DRUGS
Melting point determination: Melting point of Docusate Sodium was checked by open capillary method using digital melting point apparatus. In this apparatus drugs were melting at specific temperature. First fill the small amount of drug in capillary. Put in the apparatus. Drugs were melting at specific temperature. Record the temperature range at which it melted. Calcium Dobesilate was checked by Thiele tube which filled by silicone oil. Melting point of CAD and DOS given in table no.10
FT-IR Spectroscopy: Fourier transform infrared spectroscopy was carried out for solid samples to identify the drug. The samples were prepared by the potassium bromide disc method. Powders (1mg drug in 100mg KBr) were triturated in agate mortar and pestle to produce fine and uniform mixture. The pellets were prepared by compressing the powders at 20 psi for 10 min using Potassium bromide – press. Pure potassium bromide powder was used as background, and for baseline correction. Prepared sample disc was placed in a sample holder and transferred to sample compartment. Samples were scanned in the region of 3500-500 cm-1 using a FTIR. IR spectrum of Calcium Dobesilate and Docusate Sodium were shown in fig.1 and 2 respectively. Interpretations of spectra were shown in table 11 and 12 respectively.
Solubility: Solubility was determined by taking 10 mg of drug in 100 ml volumetric flask and adding drop by drop (0.1ml) solvents and shaking for few minutes. Solvent was added until drug dissolved completely. Solubility was measured by the required solvent quantity. Solubility data for each component are listed in table no.7. The same method was carried for determination of drug Solubility in Water, Methanol, Acetonitrile, 0.1 N NaOH and 0.1 N HCl. Data for CAD and DOS are given in table no.13.
TABLE 7: IP’2014 DATA OF SOLUBILITY STUDY
Descriptive term |
Approximate Volume of solvent in millilitres per gram of solute |
Very soluble |
Less than 1 |
Freely soluble |
From 1 to 10 |
Soluble |
From 10 to 30 |
Sparingly soluble |
From 30 to 100 |
S1lightly soluble |
From 100 to 1000 |
Very slightly soluble |
From 1000 to 10,000 |
Practically insoluble |
More than 10,000 |
UV VISIBLE SPECTOSCOPIC METHOD
Preparation of Solution: Prepare 100µg/ml and 50µg/ml of Calcium Dobesilate and 20µg/ml and 10µg/ml Docusate Sodium in different diluents of water, methanol, 0.1N HCl, 0.1N NaOH, acetonitrile, acetonitrile: water (50:50) and scanned over the range of 190 to 400nm in UV-Visible Spectrophotometer. Trials of spectra were shown in figure no.3, 4, 5, 6, 7, and 8.
RP-HPLC METHOD FOR CALCIUM DOBESILATE AND DOCUSATE SODIUM IN PHARMACEUTICAL DOSAGE FORM
Selection of wavelength: Standard solutions of Calcium Dobesilate (CAD) and Docusate Sodium (DOS) in Acetonitrile: water (50:50) was scanned over the range of 190 to 400nm in UV Visible spectrophotometer. Best response of two drugs achieved at 214nm. So, both drugs are detected at 214nm wavelength.
Selection of mobile phase: In reverse-phase chromatography, the stationary phase is less polar than the mobile phase. The mobile phase was selected in such a way to achieve the proper peak shape with sufficient height, theoretical plates, resolution and purity. Selection of mobile phase involves the selection of buffer, pH of the buffer, buffer concentration, selection of solvent and buffer to solvent ratio. The mobile phase used for screening is depicted in table 8.
Preparation of solution: Prepare 100µg/mL solution for Calcium Dobesilate and 20 µg/mL solution for Docusate Sodium.
Screening of mobile phase
TABLE8: SCREENING OF MOBILE PHASE
SR. NO |
CONDITION |
FIG. NO. |
1. |
Acetonitrile: Methanol: 0.02M Potassium dihydrogen phosphate pH-4 (25:10:65,v/v/v) |
9 |
2. |
Methanol: 0.02M Potassium dihydrogen phosphate pH-4 (30:70,v/v) |
10 |
3. |
Methanol: 0.02M Potassium dihydrogen phosphate pH-4.8 (30:70,v/v) |
11 |
4. |
Acetonitrile: 0.02M Potassium dihydrogen phosphate pH-4.8 (50:50,v/v) |
12 |
5. |
Acetonitrile: 0.02M Potassium dihydrogen phosphate pH-4.8 (70:30,v/v) |
13 |
6. |
Acetonitrile: 7mM Ammonium acetate pH-4.8(50:50,v/v) |
14 |
7. |
Acetonitrile:0.01M Tetrabutylammonium dihydrogen phosphate pH-4.8(66:34,v/v) |
15 |
Screening of Chromatographic condition for DOE (QbD approach): Stationary phase: C18 (250mm X 4.6mm), 5μm particle size.
Mobile Phase: Acetonitrile: 0.01M Tetrabutylammonium dihydrogen phosphate(pH-4.8) (66: 34, v/v) Flow Rate: 1.0 mL/min
Detection Wavelength: 214nm
QbD approach on Screening of mobile phase with Statistical tool DOE by using Box-Behnken design
· Box-Behnken is statistical technique for development and optimization of the best mobile phase in RP-HPLC.
· Design Export 9.0.3.1 software was used for the determination of optimized conditions.
· Box-Behnken design was selected and used to optimization of mobile phase.
· 3 independent factor (pH, Flow rate, Mobile phase ratio) and response (dependent factor)- (Resolution, Asymmetry factor and theoretical plates of CAD and DOS) were used in design.
· 3 Level which are using for the particular factor which are assigned as +1,0, -1 and its value are given into table.
TABLE 9: factor of 3 level and its value
LEVEL |
FLOW RATE |
pH |
MOBILE PHASE RATIO |
+1 |
0.8 |
4.6 |
60:40 (40) |
0 |
1 |
4.8 |
66:34 (34) |
-1 |
1.2 |
5.0 |
72:28 (28) |
Each 15 set of experimental run was performed, suggested by Box-Behnken design was shown in table no.14.
Optimization of Chromatographic condition for method development by Box-Behnken Design
Stationary phase: C18 (250mm X 4.6mm), 5μm particle size.
Mobile Phase: Acetonitrile: 0.01M Tetrabutylammonium dihydrogen phosphate(pH-4.8) (66: 34, v/v) Flow Rate: 1.0 mL/min. Detection Wavelength: 214nm Injection volume: 20 µl Run Time: 5 min.
Preparation of mobile phase: Based on the literature review and solubility study, Calcium Dobesilate (CAD) is freely soluble in Acetonitrile and Docusate Sodium (DOS) is also freely soluble in Acetonitrile. So, Acetonitrile: 0.01M Tetrabutylammonium dihydrogen phosphate (66:34) were selected as solvent for the method development as per USP method.
Preparation of diluent: Based on the literature review and solubility study, Calcium Dobesilate (CAD) is freely soluble in Acetonitrile and water. Docusate Sodium (DOS) is also freely soluble in Acetonitrile and soluble in water.so, for HPLC method development for as per USP method Acetonitrile: water (50:50) was selected as diluent for solution preparation.
Preparation of solutions
Stock solution of Calcium Dobesilate: Accurately weighed 50mg Calcium Dobesilate (CAD) dissolve in 50ml of diluent which contain Acetonitrile: water (50:50) in 50 ml of calibrated volumetric flask to prepare 1000µg/ml.
Stock solution of Docusate Sodium: Accurately weighed 10mg Docusate Sodium (DOS) dissolve in 50ml of diluent which contain Acetonitrile: water (50:50) in 50 ml of calibrated volumetric flask to prepare 200µg/ml.
Preparation of Calcium Dobesilate and Docusate Sodium standard solution: Aliquot of 25 ml from Calcium Dobesilate (CAD) 1000µg/ml stock solution and 25ml from Docusate Sodium (DOS) 200µg/ml stock solution in 50ml calibrated volumetric flask and made up to mark with of diluent which contain Acetonitrile: water (50:50) to prepare 500µg/ml Calcium Dobesilate (CAD) and 100µg/ml Docusate Sodium (DOS).
Preparation of working standard solutions: Aliquots of 1, 3, 5, 7, 9ml from above CAD and DOS stock solution were separately transferred in to 10ml of calibrated volumetric flask and made up to mark to produce solutions having concentration of Calcium Dobesilate 50, 150, 250, 350, 450µg/ml and Docusate Sodium 10, 30, 50, 70, 90µg/ml with diluent which contain Acetonitrile: water (50:50) in Standard solutions were filtered through 0.22μm nylon membrane filter paper and first few drops of filtrate were discarded.
METHOD VALIDATION
Validation of developed method was done as per ICH Q2 (R1) guidelines. According to ICH guidelines validation parameters to be performed for a newly developed
RPHPLC method are:
· System suitability
· Specificity
· Linearity
· Range
· Precision
o Repeatability
o Intermediate Precision(Reproducibility)
o Inter day Precision
· Accuracy
· Limit of Detection (LOD)
· Limit of Quantitation (LOQ)
· Robustness:
o Different Flow rate
o Different Wavelength
o Different mobile phase ratio
System Suitability Testing: System suitability testing is an integral part of analytical procedures. It can be defined as tests to ensure that the method can generate results of acceptable accuracy and precision. The USP defines parameters that can be used to determine system suitability prior to analysis. These parameters include plate number (N), tailing factor (Tf), resolution (Rs), and relative standard deviation (%RSD) of peak height or peak area for repetitive injections (n=6). At least two of these criteria are required to demonstrate system suitability for any method. Data of system suitability is given in table no.5.11.
Specificity: Specificity is achieved by observing blank and that product excipients peak was not interfering with the sample peak during the validation study. Result of specificity is given in figure no.28, 29, and 30.
Linearity: This linearity studies be performed by using at least five different concentration levels. The calibration curves were plotted over a wide concentration range and the linear response was observed over a range of 50-450μg/ml for Calcium Dobesilate and 10-90μg/ml for Docusate Sodium. Aliquots of 1, 3, 5, 7 and 9mL was accurately pipetted out from standard stock solution (500μg/ml CAD and 100μg/ml DOS) of and then transferred to a series of 10 ml of calibrated volumetric flasks and make up to the mark with HPLC-grade Acetonitrile: Water (50:50) diluent. Repeating this process at least three times with different stock solutions generates three complete sets of linearity. Calibration curves were constructed by plotting peak areas vs. concentration of drug, and the regression equations were calculated. 20μl of each solution were injected under the optimised operating chromatographic conditions. Data of linearity is given in 31, 32 figure no. table no.16. Acceptance limit: Correlation Coefficient R2 > 0.999.
Range: The Range of a method can be defined as the Lower and Upper concentrations for which the analytical method has adequate accuracy, precision, and linearity. Based on the linearity study, linear response was observed over a range of 50-450μg/ml for Calcium Dobesilate and 10-90μg/ml for Docusate Sodium simultaneously.
Precision
Repeatability: The precision of the instrument was checked by repeatedly injecting (n = 6) standard solutions 250μg/ml of CAD and 50μg/ml of DOS simultaneously under the same chromatographic condition and peak area, retention time and tailing factor was measured at 214nm. Data of repeatability is given in table no.17. Acceptance criteria: % RSD should be NMT 2 %
Intermediate precision: Intraday Precision Three replicates of three concentrations containing mixture of CAD and DOS (150, 250, 350μg/ml of CAD and 30, 50, 70μg/ml of DOS) total nine determinations were analysed at three different time in one day and chromatogram was recorded at 214.0nm. %RSD was calculated. Data of Intraday precision is given in table no.18. Acceptance criteria: %RSD should be NMT 2%.
Interday Precision: Three replicates of three concentrations containing mixture of CAD and DOS (150, 250, 350μg/ml of CAD and 30, 50, 70μg/ml of DOS) total nine determinations were analysed at three consecutive day and chromatogram was recorded at 214.0nm. %RSD was calculated. Data of Interday precision is given in table no.19. Acceptance criteria: %RSD should be NMT 2%.
Accuracy (%Recovery): The accuracy of a measurement is defined as the closeness of the measured value to the true value. Accuracy is represented and to determine an analyte in formulation by a spiked recovery studies. The recovery at each level was determined by comparison of the samples spiked with known amounts of the analyte. For assay, three spiked levels 80%, 100%, 120% were selected. Known concentration (CAD 350g/ml, DOS 70μg/ml) of formulation spiked with (CAD 280, 350, 420μg/ml, DOS 56, 70, 84μg/ml) solution of API to generate the three replicate of each level (CAD 280μg/ml (80%), 350μg/ml (100%), and 420μg/ml (120%), DOS 56μg/ml (80%), 70μg/ml (100%), and 84μg/ml (120%)). Each level was measured and % recovery was calculated from % Recovery = (Experimental value / Actual value) × 100 Data of Accuracy is given in table no.20.
Limit of Detection and Quantification LOD: The LOD was carried out based on the Standard Deviation of the Response and the Slope. The LOD was estimated from the set of five calibration curves used to determine method linearity. The LOD may be calculated as, LOD = 3.3 × (SD/ Slope). Where, SD = Standard deviation of the Y- intercepts of the five calibration curves, Slope (S) = Mean slope of the five calibration curves. LOQ: It was carried out based on the Standard Deviation of the Response and the Slope. The LOQ was estimated from the set of five calibration curves used to determine method linearity. The LOQ may be calculated as, LOQ = 10 × (SD /Slope). Where, SD = Standard deviation of the Y- intercepts of the five calibration curves. Slope(S) = Mean slope of the five calibration curves. Data of LOD and LOQ are given in table no.21.
Robustness: The robustness of analytical method is a measure of its capacity to remain unaffected by small, but deliberate variations in the method parameters and provides an indication of its reliability during normal usage. To confirm robustness change was done in flow rate (± 2 ml/min), wavelength (± 2nm) and mobile phase ratio. %RSD for area was calculated which should be NMT 2%. Data of Robustness parameter are given in table no.22, 23, 24.
Applicability of method in formulation: The response of solution of Capsule dosage form was measured at 214 nm under the chromatographic condition mentioned earlier for the quantitation of CAD and DOS. The amounts of CAD and DOS present in solution of capsule dosage form were determined by applying values of the peak area to the regression equations of the calibration graph. Data of analysis of Drugs in Capsule dosage form were shown in table no.25.
RESULT AND DISCUSSION:
IDENTIFICATION TESTS
Melting point determination
TABLE 10: MELTING POINT OF CALCIUM DOBESILATE AND DOCUSATE SODIUM
|
Reported M.P.19 |
Observed M.P. |
CALCIUM DOBESILATE |
>300°C |
>300°C |
DOCUSATE SODIUM |
153-157°C |
156-160°C |
FTIR (Fourier transform infrared spectroscopy)
A) FTIR SPECTRA OF CALCIUM DOBESILATE
Figure 1: FTIR spectra of Calcium Dobesilate
TABLE 11: INTERPRETATION OF CALCIUM DOBESILATE
FUNCTIONAL GROUP |
CHARACTERISTIC ABSORPTION -1 (cm) |
S-O |
1081 |
C-S |
1211 |
S=O |
1022 |
S=O |
1163 |
OH |
3430 |
CH Aromatic |
3152(stretching) |
C=C aromatic |
1511,1654 |
Figure 2: FTIR spectra of Docusate Sodium
Table 12: Interpretation of Docusate Sodium
FUNCTIONAL GROUP |
CHARACTERISTIC ABSORPTION (cm-1) |
C-H Aliphatic |
2961.81 |
C-O |
1044 |
C=O |
1738.05 |
OH |
2933.92 |
Solubility
Table 13: Solubility of Calcium Dobesilate and Docusate Sodium
Solvents |
Calcium Dobesilate |
Docusate Sodium |
Water |
Very soluble |
Sparingly Soluble |
Methanol |
Soluble |
Freely soluble |
Acetonitrile |
Freely soluble |
Soluble |
0.1 N NaOH |
Freely soluble |
slightly soluble |
0.1 N HCl |
Freely soluble |
slightly soluble |
UV-VISIBLE SPECTROSCOPIC METHOD DEVELOPMENT
Different trials for selection of solvent
DEVELOPMENT OF HPLC METHOD
HPLC METHOD DEVELOPMENT TRIALS
SCREENING METHOD APPLIED FOR DOE APPROACH
DESIGN OF EXPERIMENT
Here BOX BEHNKEN Design is applied for developing Quality in method. Box-Behnken design with 15 runs
Box-Behnken Design
TABLE 14: Box-Behnken Design
STD |
RUN |
FACTOR1 FLOW RATE |
FACTOR2 pH |
FACTOR3 MPR |
RESO LUTION |
ASYM METRY (CD) |
ASYMMETRY (DS) |
TP (CD) |
TP (DS) |
7 |
1 |
0.8 |
4.8 |
40 |
4.813 |
1.849 |
1.526 |
1159 |
1079 |
10 |
2 |
1 |
5 |
28 |
5.268 |
2.147 |
1.886 |
1437 |
1196 |
6 |
3 |
1.2 |
4.8 |
28 |
5.684 |
1.956 |
1.571 |
1497 |
1256 |
14 |
4 |
1 |
4.8 |
34 |
5.900 |
1.218 |
1.213 |
5117 |
4931 |
5 |
5 |
0.8 |
4.8 |
28 |
5.319 |
1.996 |
1.583 |
2098 |
1831 |
2 |
6 |
1.2 |
4.6 |
34 |
5.814 |
1.935 |
1.642 |
2171 |
1817 |
11 |
7 |
1 |
4.6 |
40 |
4.725 |
1.994 |
1.785 |
1449 |
1184 |
4 |
8 |
1.2 |
5 |
34 |
5.317 |
2.156 |
1.865 |
1836 |
1523 |
12 |
9 |
1 |
5 |
40 |
4.743 |
1.954 |
1.763 |
1350 |
1097 |
13 |
10 |
1 |
4.8 |
34 |
5.807 |
1.274 |
1.234 |
5094 |
4895 |
8 |
11 |
1.2 |
4.8 |
40 |
4.569 |
1.901 |
1.521 |
2135 |
1756 |
1 |
12 |
0.8 |
4.6 |
34 |
5.429 |
1.967 |
1.665 |
1899 |
1660 |
9 |
13 |
1 |
4.6 |
28 |
5.723 |
1.996 |
1.693 |
1684 |
1540 |
3 |
14 |
0.8 |
5 |
34 |
5.294 |
1.994 |
1.847 |
1821 |
1568 |
15 |
15 |
1 |
4.8 |
34 |
5.884 |
1.205 |
1.200 |
5164 |
4915 |
DATA OF RESOLUTION
FIGURE 16: COUNTOUR PLOT OF RESOLUTION ON pH, MOBILE PHASE RATIO AND FLOW RATE
FIGURE 17: 3D SURFACE PLOT OF RESOLUTION ON pH, MOBILE PHASE RATIO AND FLOW RATE
FULL MODEL POLYNOMIAL EQUATION: Resolution= +5.86+0.066*A-0.13*B-0.39*C-0.090*AB-0.15*AC+0.12*BC-0.21A2-0.19B2-0.56C2
DATA OF ASYMMETRY OF CALCIUM DOBESILATE (CAD)
ANOVA OF ASYMMETRY OF CALCIUM DOBESILATE
FIGURE 18: COUNTOUR PLOT OF ASYMMRTEY OF CALCIUM DOBESILATE ON pH, MOBILE PHASE RATIO AND FLOW RATE
FIGURE 19: 3D SURFACE PLOT OF ASYMMRTEY OF CALCIUM DOBESILATE ON pH, MOBILE PHASE RATIO AND FLOW RATE
FULL MODEL POLYNOMIAL EQUATION:
Asymmetry (CAD)=+1.23+0.017*A+0.045*B-0.050*C+0.049*AB+0.024*AC 0.048*BC+0.034*A2+0.44*B2+0.35*C2
POLYNOMIAL REGRESSION EQUATION: +1.23+0.017*A+0.045*B 0.050*C+0.049*AB+0.048*BC+0.034*A2+0.44*B2+0.35*C2
DATA OF ASYMMETRY OF DOCUSATE SODIUM (DOS)
ANOVA OF ASYMMETRY OF DOCUSATE SODIUM
FIGURE 20: COUNTOUR PLOT OF ASYMMRTEY OF DOCUSATE SODIUM ON pH, MOBILE PHASE RATIO AND FLOW RATE
FIGURE 21: 3D SURFACE PLOT OF ASYMMRTEY OF DOCUSATE SODIUM ON pH, MOBILE PHASE RATIO AND FLOW RATE
FULL MODEL POLYNOMIAL EQUATION:
ASYMMETRY (DOS) = +1.22-2.73*10-3*A+0.072*B-0.017*C+0.010*AB+1.7*103AC 0.054*BC+0.15A2+0.39*B2+0.18*C2
POLYNOMIAL REGRESSION EQUATION: +1.22+0.072*B-0.054*BC+0.15A2+0.39*B2+0.18*C2
5.3.3.4 DATA OF THEORETICAL PLATES OF CALCIUM DOBESILATE (CAD)
ANOVA OF THEORETICAL PLATES OF CALCIUM DOBESILATE
FIGURE 22: COUNTOUR PLOT OF THEORETICAL PLATES OF CALCIUM DOBESILATE ON pH, MOBILE PHASE RATIO AND FLOW RATE
FIGURE 23: 3D SURFACE PLOT OF THEORETICAL PLATES OF CALCIUM DOBESILATE ON pH, MOBILE PHASE RATIO AND FLOW RATE
FULL MODEL POLYNOMIAL EQUATION:
Theoretical plates (CAD)= +5125.0+82.75*A-94.88*B-77.88*C-64.25*AB+394.25*AC+37.00*BC-1475.50*A2-1717.75*B2-1929.25*C2
DATA OF THEORETICAL PLATES OF DOCUSATE SODIUM (DS)
ANOVA OF THEORETICAL PLATES OF DOCUSATE SODIUM
24: COUNTOUR PLOT OF THEORETICAL PLATES OF DOCUSATE SODIUM ON pH, MOBILE PHASE RATIO AND FLOW RATE
FIGURE
FIGURE 25: 3D SURFACE PLOT OF THEORETICAL PLATES OF DOCUSATE SODIUM ON pH, MOBILE PHASE RATIO AND FLOW
RATE FULL MODEL POLYNOMIAL EQUATION:
Theoretical plates (DOS) = +4913.67+39.37*A-114.75*B-88.38*C-75.75*AB+313.00*AC+64.25*BC-1510.08*A2-1736.33*B2-1923.08*C2
FIGURE 26: OVERLAY PLOT FOR OPTIMIZATION OF HPLC METHOD
OPTIMIZED CONDITION FROM DESIGN OF EXPERIMENT
Mobile Phase: Acetonitrile: 0.01 M Tetrabutylammonium hydrogen phosphate(pH-4.8) (66: 34, v/v)
Flow Rate: 1.0 ml/min Wavelength: 214nm
Discussion: from the overlay plot it is concluding that Method development with QbD by statistical tool DOE gives best optimization condition for validation of the method.
METHOD VALIDATION
System suitability testing
Figure 27: System suitability testing
Table 15: Data of system suitability
Sr. No. |
Retention time(min) |
Theoretical plates |
Tailing factor |
Resolution |
|||
CAD |
DOS |
CAD |
DOS |
CAD |
DOS |
||
1. |
2.943 |
4.193 |
4218 |
4970 |
1.263 |
1.386 |
5.964 |
2. |
2.940 |
4.183 |
3957 |
4507 |
1.235 |
1.354 |
5.701 |
3. |
2.933 |
4.170 |
3940 |
4915 |
1.238 |
1.326 |
5.882 |
4. |
2.933 |
4.167 |
3940 |
4682 |
1.365 |
1.374 |
5.729 |
5. |
2.940 |
4.183 |
3957 |
4719 |
1.240 |
1.362 |
5.776 |
6. |
2.943 |
4.170 |
3950 |
4842 |
1.239 |
1.375 |
5.753 |
LIMIT |
|
>2000 |
<2 |
>2 |
Discussion
· Theoretical plate count of Calcium Dobesilate and Docusate Sodium is greater than 2000.
· The tailing factor of six replicate of Calcium Dobesilate and Docusate Sodium is less than 2.0.
Resolution of the peak is greater than 2.0.
Specificity
Figure 28: Chromatogram of Blank
Figure 29: Chromatogram of standard solution of CAD and DOS
Figure 30: Chromatogram of Capsule dosage form
DISCUSSION:
By observing chromatogram of blank, standard solution of CAD and DOS, and product it can be seen that there was no interference of excipients during the validation study.
Linearity and Range
The linearity is 50-450 µg/ml for Calcium Dobesilate and 10-90 µg/ml for Docusate Sodium. Linearity and range data was depicted in figure 22 and table 12.
Figure 31: Linearity of CAD and DOS in mixture
Figure 32: Linearity of CAD
Figure 33: Linearity of DOS
Table 16: Data of Linearity and Range
Sr. No. |
CAD |
|
DOS |
|
Conc. (µg/ml) |
Mean Peak Area (mV.s) ±SD |
Conc. (µg/ ml) |
Mean Peak Area (mV.s) ±SD |
|
1 |
50 |
1662.518±0.192 |
10 |
197.794±0.094 |
2 |
150 |
1995.359±0.141 |
30 |
232.616±0.153 |
3 |
250 |
2310.433±0.096 |
50 |
264.129±0.149 |
4 |
350 |
2658.089±0.157 |
70 |
296.212±0.097 |
5 |
450 |
2971.491±0.094 |
90 |
328.047±0.173 |
Figure 34: Calibration curve for Calcium Dobesilate
Figure 35: Calibration curve for Docusate Sodium
Discussion: For CAD, regression equation was found to be y = 3.2807x + 1499.4 and correlation co-efficient (R2) was found to be 0.9998.
For DOS, regression equation was found to be y = 1.6205x + 182.73 and correlation co-efficient (R2) was found to be 0.9997.
Precision
A) Repeatability
Table 17: Data of Repeatability
Calcium Dobesilate |
Docusate Sodium |
|
|
Concentration (µg/ml) (n=6) |
Peak Area (mV.s) |
Concentration (µg/ml)(n=6) |
Peak Area (mV.s) |
250 |
2235.694 |
50 |
254.395 |
250 |
2268.562 |
50 |
251.257 |
250 |
2274.407 |
50 |
249.305 |
250 |
2238.269 |
50 |
244.512 |
250 |
2242.486 |
50 |
250.609 |
250 |
2278.328 |
50 |
245.685 |
Mean(n=6) |
2256.291 |
Mean(n=6) |
249.293 |
S.D. |
19.514 |
S.D. |
3.674 |
%RSD |
0.864 |
%RSD |
1.473 |
Discussion: The % RSD for CAD and DOS was found to be 0.864 and 1.473 respectively. Limit: % RSD should be < 2.
B) Intermediate precision Intraday Precision
Table 18: Data of Intraday precision
Drug |
Conc. (µg/ml) |
Peak Area(mV.s) |
|
Mean Area(vs.)(n=3) |
SD |
%RSD |
|
Set 1 |
Set 2 |
Set 3 |
|||||
CAD |
150 |
1954.249 |
1975.394 |
1966.570 |
1965.404 |
10.620 |
0.540 |
250 |
2256.357 |
2274.407 |
2242.486 |
2257.750 |
16.006 |
0.708 |
|
350 |
2658.214 |
2621.014 |
2656.052 |
2645.093 |
20.881 |
0.789 |
|
DOS |
30 |
201.354 |
197.794 |
198.885 |
199.344 |
1.823 |
0.914 |
50 |
253.598 |
257.248 |
250.609 |
253.818 |
3.324 |
1.309 |
|
70 |
269.953 |
265.981 |
263.214 |
266.382 |
3.387 |
1.271 |
Discussion: The % RSD was found to be 0.540-0789 and 0.914-1.309 for CAD and DOS respectively.
C) Inter Day Precision
Table 19: Data of Interday precision
Drug |
Conc. (µg/ml) |
Peak Area(mV.s) |
Mean Area (mV.s) (n=3) |
SD |
%RSD |
||
Day 1 |
Day 2 |
Day 3 |
|||||
CAD |
150 |
1960.345 |
1978.489 |
1980.215 |
1973.016 |
11.007 |
0.557 |
250 |
2253.987 |
2286.384 |
2256.357 |
2265.576 |
18.059 |
0.797 |
|
350 |
2708.098 |
2756.173 |
2743.718 |
2735.996 |
24.950 |
0.911 |
|
DOS |
30 |
215.814 |
220.109 |
218.918 |
218.280 |
2.217 |
1.015 |
50 |
260.287 |
265.761 |
266.990 |
264.346 |
3.568 |
1.349 |
|
70 |
276.181 |
270.791 |
279.321 |
275.431 |
4.314 |
1.566 |
Discussion: The % RSD was found to be 0.557-0.911 and 1.015-1.566 for CAD and DOS respectively.
Accuracy
Table 20: Data of accuracy
Drug |
Level |
Amount of Sample (µg/ml) |
Amount of Standard Spiked(µg/ml) |
Total Amount (µg/ml) |
Amount Recovery (µg/ml) |
%Recovery |
CAD |
80% |
150 |
120 |
270 |
268.519 |
99.451 |
100% |
150 |
150 |
300 |
300.247 |
100.082 |
|
120% |
150 |
180 |
330 |
329.985 |
99.995 |
|
DOS |
80% |
30 |
24 |
54 |
54.641 |
101.187 |
100% |
30 |
30 |
60 |
58.843 |
98.071 |
|
120% |
30 |
36 |
66 |
66.673 |
101.019 |
Discussion: Accuracy was found to be 99.451% - 100.082% and 98.071% - 101.187% for CAD and DOS respectively.
LOD and LOQ
Table 21: Data of LOD and LOQ
PARAMETER |
CAD |
DOS |
S.D. of the Y- Intercepts of the 5 calibration curves |
3.640 |
1.897 |
Mean slope of the 5 calibration curves |
3.216 |
1.574 |
LOD = 3.3 × (SD/Slope) |
3.734 |
3.977 |
LOQ = 10 × (SD/Slope) |
11.316 |
12.053 |
Discussion: The LOD and LOQ for CAD were found to be 3.734 and 11.316 respectively. The LOD and LOQ for DOS were found to be 3.977 and 12.053 respectively.
Robustness:
The change was done in flow rate (±0.2 ml/min), wavelength (±2 nm) and mobile phase ratio. %RSD for area was calculated which should be less than 2%. The data were shown in table no. 5.34, 5.35, and 5.36 respectively.
Table 22: Data for flow rate change
Drug |
Conc. (µg/ml) |
Area(mV.s) |
Mean Area (mV.s) |
SD |
%RSD |
||
0.8ml/min |
1.0ml/min |
1.2ml/min |
|||||
CAD |
150 |
2355.606 |
2367.248 |
2348.367 |
2357.074 |
9.525 |
0.404 |
250 |
2868.285 |
2845.217 |
2865.965 |
2859.822 |
12.701 |
0.444 |
|
350 |
3339.495 |
3326.946 |
3354.457 |
3340.299 |
13.773 |
0.412 |
|
DOS |
30 |
203.459 |
204.675 |
207.549 |
205.227 |
2.100 |
1.023 |
50 |
283.295 |
28.397 |
286.217 |
283.636 |
2.428 |
0.856 |
|
70 |
304.902 |
309.212 |
305.38 |
306.499 |
2.361 |
0.770 |
The % RSD was found to be 0.404-0.444 and 0.856-1.023 for CAD and DOS respectively.
Table 23: Data for wavelength change
Drug |
Conc.(µg/ml) |
Area(mV.s) |
Mean Area (mV.s) |
SD |
%RSD |
||
212nm |
214nm |
216nm |
|||||
CAD |
150 |
1905.943 |
1854.331 |
1859.328 |
1873.201 |
28.456 |
1.519 |
250 |
2295.395 |
2310.433 |
2261.376 |
2289.068 |
25.133 |
1.097 |
|
350 |
2815.374 |
2739.143 |
2749.617 |
2768.045 |
41.321 |
1.492 |
|
DOS |
30 |
205.607 |
204.675 |
206.991 |
205.757 |
1.165 |
0.566 |
50 |
224.816 |
229.374 |
231.265 |
228.485 |
3.315 |
1.450 |
|
70 |
295.328 |
299.853 |
291.349 |
295.510 |
4.254 |
1.439 |
Discussion: The % RSD was found to be 1.097-1.519 and 0.566-1.450 for CAD and DOS respectively.
Table 24: Data for mobile phase ratio change
Drug |
Conc.(µg/ml) |
Peak Area (mV.s) |
Mean Area(mV.s) |
SD |
%RSD |
||
64:36(v/v) |
66:34(v/v) |
68:32(v/v) |
|||||
CAD |
150 |
1994.146 |
1992.429 |
2017.392 |
2001.322 |
13.943 |
0.696 |
250 |
2369.429 |
2315.391 |
2367.253 |
2350.691 |
30.590 |
1.301 |
|
350 |
2739.252 |
2698.982 |
2743.952 |
2727.186 |
24.718 |
0.906 |
|
DOS
|
30 |
197.721 |
202.641 |
198.372 |
199.578 |
2.672 |
1.339 |
50 |
230.397 |
235.943 |
231.641 |
232.660 |
2.910 |
1.250 |
|
70 |
265.742 |
271.429 |
269.571 |
268.914 |
2.899 |
1.078 |
The % RSD was found to be 0.696-1.301 and 1.078-1.339 for CAD and DOS respectively.
Applicability of method in Capsule dosage form
Figure 36: Chromatogram of Drugs in capsule dosage form
Table 25: Data of analysis of Drugs in capsule dosage form
Sample No. |
Label Claim(mg) |
Amount Found (mg) |
% Label Claim
|
|||
CAD |
DOS |
CAD±SD |
DOS±SD |
MAS±SD |
DOS±SD |
|
1 |
500 |
100 |
498.948 |
99.159 |
99.789 |
99.159 |
2 |
500 |
100 |
499.374 |
98.736 |
99.874 |
98.736 |
3 |
500 |
100 |
498.862 |
98.627 |
99.772 |
98.627 |
Mean (n=3) |
499.061±0.274 |
98.840±0.281 |
99.811±0.054 |
98.840±0.281 |
Discussion: % Assay CAD and DOS was found to be 99.772% - 99.874% and 98.159% - 98.736% respectively.
SUMMARY:
A specific, accurate and precise RP-HPLC method has been developed and validated for estimation of Calcium Dobesilate and Docusate Sodium in Capsule dosage form. The condition was optimized to obtain an adequate separation of the compounds. There is no any UV spectroscopy method of this combination due to very less absorbance of the Docusate Sodium in various solvent. Various trials were taken during method development for Screening of method in RP-HPLC. Apply QbD approach by using statistical tool Design of experiment (DOE) with BoxBehnken design for the optimization of the screening method of RP-HPLC method.
Calcium Dobesilate and Docusate Sodium were estimated on Phenomenex C18 (250mm X 4.6mm), 5μm column using Acetonitrile: 0.01M Tetrabutylammonium dihydrogen phosphate pH-4.8(66:34, v/v) as mobile phase with flow rate 1.0 ml/min and detection was carried out at 214 nm. The retention time of Calcium Dobesilate and Docusate Sodium were found to be 2.943 min and 4.190 min respectively. The accuracy, precision and other validation parameters were determined and validated statistically. Parameters which were validated are as follows:
Table 26: Validation Parameters
VALIDATION PARAMETERS |
Calcium Dobesilate |
Docusate Sodium |
Linearity and range |
50-450 μg/ml |
10-90μg/ml |
Co-relation coefficient (R2) |
0.9998 |
0.9997 |
LOD,LOQ |
3.734 and 11.316 |
3.977 and 12.053 |
%RSD limit of precision |
<2 |
<2 |
%RSD of Repeatability |
0.864 |
1.473 |
%RSD of Intraday precision |
0.540-0789 |
0.914-1.309 |
%RSD of Interday precision |
0.557-0.911 |
1.015-1.566 |
Accuracy |
99.451% - 100.082% |
98.071% - 101.187% |
%RSD limit of |
<2 |
<2 |
Robustness |
|
|
Change in flow rate |
0.404-0.444 |
0.856-1.023 |
Change in wavelength |
1.097-1.519 |
0.566-1.450 |
Change in Mobile phase ratio |
0.696-1.301 |
1.078-1.339 |
%recovery |
99.772% - 99.874% |
98.159% - 98.736% |
REFERENCE:
1. Kasture AV., Wadodkar SG., and Mahadik KR. Introduction to Instrumental Techniques; 12th Edn; Nirali Prakashan, Pune, 2002, pp 1-3.
2. Skoog DA., Holler FJ., and Nieman TA. An Introduction to Analytical Chemistry; 5th Edn; Thomson Brooks/Cole Publication, Singapore, 1994, pp 566-568.
3. International conference on harmonization, “Guidelines Q8 Pharmaceutical development” http//ich.org/products/guidelines/quality/article/quality- guidelines.html
4. Chatterjee S. , “QBD consideration for analytical methods-FDA perspective”, http://www.fda.gov/downloads/AboutFDA/centersOffices/OfficeofMedicalProducts and Tobacco/CDER/UCM359266.pdf
5. Indian Pharmacopoeia; 7Th Edn(2); Indian Pharmacopoeia Commission, Ghaziabad, 2014, pp 1250-1251.
6. Indian Pharmacopoeia; 7Th Edn(1); Indian Pharmacopoeia Commission, Ghaziabad, 2014, pp 183
7. Indian Pharmacopoeia; 7Th Edn(2); Indian Pharmacopoeia Commission, Ghaziabad, 2014, pp 1610-1611.
8. United State Pharmacopoeia; USP 32 NF 27(2); United State Pharmacopeia convention, 2008, pp 2006-2007.
9. “Marketed Formulation”, July 2014 http://www.medicineindia.org/medicine-brand-details/9100/smuth-capsule
10. “Marketed Formulation”, July 2014 http://www.medindia.net/drug-price/docusate-combination/swift.htm.
11. British Pharmacopoeia; medicine and health care products regulatory Agency, Volume 1, 2010, pp 327.
12. British Pharmacopoeia; medicine and health care products regulatory Agency, Volume 1, 2010, pp 710.
13. “Detection of calcium dobesilate in calcium dobesilate capsules: a comparison of HPLC and UV spectrophotometry.” J. of Medicine and Hygiene. 2009.
14. Rona K., Ary K., “Determination of calcium dobesilate in human plasma using ion- pairing extraction and high-performance liquid chromatography.” Journal of Chromatography. 2001, 755, 245-251.
15. Hepsebah NJR., Nihitha D., Padma P., “RP-HPLC method development and validation for the simultaneous quantitative estimation of Troxerutin and Calcium Dobesilate in tablet” Intenational J. Pharm. Science. 2014, 6(1), 333-339.
16. Hepsebah NJR., Nihitha D., Padma P., “RP-HPLC method development and validation for the simultaneous quantitative estimation of Troxerutin and Calcium
17. Hoque DR., Zimmardi JA., Shah KA., “HPLC analysis of docusate sodium in soft gelatin capsules” J Pharm Science. 1992, 81(4), 359-361.
18. Chandratreya P., Mahajan H., “Development of simple and rapid method for estimation of low levels of docusate sodium from equipment surfaces” International J. of chem. tech. research. 2014, 6(14), 5680-5686.
19. The Merck Index; 14th Edn; Merck Research Laboratories, 2006, pp 3406.
Received on 03.02.2021 Modified on 30.03.2021
Accepted on 20.05.2021 ©A&V Publications All right reserved
Res. J. Pharma. Dosage Forms and Tech.2021; 13(3):213-229.
DOI: 10.52711/0975-4377.2021.00038