Development and
Validation of visible Spectrophotometric method of Tapentadol
Hydrochloride in bulk and Pharmaceutical dosage form
Madhu Babu
K., Kathirvel S.*
Department
of Pharmaceutical Analysis, Hindu College of Pharmacy, Amaravathi Road,
Guntur, Andhra Pradesh – 522002
ABSTRACT:
A simple, sensitive, accurate spectrophotometric method has been
developed for the estimation of Tapentadol
hydrochloride in bulk and pharmaceutical formulation. The method is based on
the formation of blue colored chromogen with Folin-Ciocalteu reagent in the presence of NaOH and obeys beer’s law in the
range of 5 - 35µg/mL exhibiting the maximum
absorbance at 763nm. The limit of detection and quantification were found to be
0.76µg/mL and 2.33µg/mL
respectively. The calibration curve demonstrated a linear relationship between
the absorbance and concentration, with the correlation coefficient higher than
0.9999. The Regression Equation of the curve was Y=0.0248x-0.0436. The method
was fully validated according to ICH guidelines.
KEYWORDS: Tapentadol hydrochloride, Folin-Ciocalteu
(FC) reagent, NaOH, Validation, ICH guidelines.
INTRODUCTION:
Chemically Tapentadol hydrochloride1
is 3-[(1R, 2R)-3-(dimethylamino)-1-ethyl-2-methylpropyl]
phenol hydrochloride. It is a centrally acting analgesic with dual mode of
action as an agonist at the µ-opioid receptor and as
a norepinephrine reuptake inhibitor2. It
is used in the treatment of moderate to severe pain3. Analysis plays an important role in the formulation development
of any drug molecule. A suitable and validated method has to be
available for the analysis of drugs in bulk, in drug delivery systems, in
dissolution studies (in vitro), and in biological samples (in vivo). If such a
suitable method for a specific need is not available, then it becomes essential
to develop a simple, sensitive, accurate, precise and reproducible method for
the estimation of drug samples. The literature survey reveals that Tapentadol was analyzed in urine sample by the Ultra
Performance Liquid Chromatography method4. More over Visible
Spectrophotometric method has not been reported in bulk and pharmaceutical
formulation. Thus the present study was undertaken to develop a colorimetric
method of Tapentadol by using FC reagent5,
and to validate the method as per ICH guidelines6-8.
MATERIALS AND METHODS:
An UV-Visible double beam spectrophotometer of Elico,
Model: SL-244 with Spectral band width of 2 nm and a pair of 1 cm matched
quartz cell was used. A pure drug sample of Tapentadol
was supplied by Lupin Pharmaceuticals as gift
samples. Tablets of Tapentadol i.e. Tydol (Ranbaxy Laboratories Ltd) was procured from the
local market.
Chemicals and Reagents:
0.1N Hydrochloric acid, Folin–Ciocalteu
Reagent (FCR) 2N (1:2 Dilution with water), 5% NaOH
(w/v) solution (5% solution of sodium hydroxide was prepared by dissolving 5g
of NaOH in sufficient quantity of water and finally
diluting to 100 ml with water).
Figure.1. Structure of Tapentadol Hydrochloride
Optimization of parameters
for colorimetric method:
All the optimization parameters are estimated at room temperature.
Tapentadol
was found to yield
a blue-colored product with
FCR and NaOH and
has absorbance maxima (λmax) at 763 nm. Therefore,
investigations were carried out to establish the most favorable conditions for
the formation of this colored product. The influence of the concentration as
well as volume of reagent on the reaction has been studied. Different concentrations and
different volumes were
tried for all the
reagents, by varying
one parameter at
a time. The optimum concentration of FCR was 2N and
of NaOH was 5 %, similarly optimum volume of FCR and NaOH was found to be 1 ml and 1 ml respectively. The
optimum concentration and volume were selected on the basis of their ability to
give maximum absorbance. Also the color
stability of newly
formed complex was measured
and the color
was found to
be stable for a
period of one
hour.
Procedure:
Preparation of the standard stock solution:
A stock solution (1000 µg/mL) of Tapentadol hydrochloride was prepared by dissolving 100 mg
of drug in 100 mL of 0.1N HCl
(stock solution-I). From the above solution 10 mL was
again diluted to 100 mL with 0.1N HCl
to get 100 µg/mL solution of tapentadol
HCl (stock solution-II). Suitable aliquots of the
drug solution (0.5 to 3.5 ml) were taken in a series of 10 ml volumetric
flasks. To each flask was added 1ml of 5% NaOH
solution. All the flasks were shaken well for at least 3 to 5 min. followed by
the addition of 1 ml of FCR solution. Finally volume was made up to the mark
with 0.1N HCl to prepare a series of standard
solutions containing 5-35 µg/mL tapentadol
HCl. The absorbance of blue colored chromogen was measured at 763 nm against reagent blank
within one hour and the absorbance were plotted against the respective
concentrations to obtain the calibration curve.
Preparation of sample solution:
The proposed method was applied to analyze the commercially available Tapentadol hydrochloride tablets (Tydol,
100mg Ranbaxy Laboratories Ltd). Ten tablets were weighed and powdered. The
amount of tablet powder equivalent to 100 mg of tapentadol
was weighed accurately and transferred to a 100 mL
volumetric flask containing 10 ml of 0.1N HCl, and sonicated for 20 min for complete dissolution of tapentadol HCl and then the
sample solution was filtered and diluted to 100 ml with 0.1N HCl to get concentration of 1000 µg/mL.
Appropriate aliquots were then taken in such a way that to produce a final
concentration of 10 µg/mL.
Method Validation:
Validation is a process of establishing documented evidence, which
provides a high degree assurance that a specific activity will consistently
produce a desired result, or a product meeting its predetermined specifications
and quality characteristics. The method was validated for different parameters
like Linearity, Accuracy, Precision, Specificity, Ruggedness, Limit of
detection (LOD) and Limit of quantification (LOQ).
Linearity:
Various aliquots were prepared from the stock solution-II (100µg/mL) ranging from 5-35 µg/mL. The
samples were analyzed with the help of a UV-VIS Spectrophotometer using 0.1N HCl as used as blank. The correlation coefficient of the
linearity was found to be 0.999 respectively.
Accuracy:
To check the accuracy of the developed methods and to study the
interference of formulation excipients, analytical
recovery studies were carried out by preparing solutions of different
concentrations, i.e., 80, 100, 120% in which the amount of marketed formulation
was kept constant and the amount of pure drug was varied respectively. The
solutions were prepared in triplicate and the accuracy was indicated by % Recovery.
Precision:
The precision of the method was demonstrated by intra-day and inter-day
variation studies. In the inter-day variation study, the solutions of same
concentration were prepared and analyzed thrice, for three consecutive days,
and the absorbances were recorded. In the intra-day
variation study, the solutions of the same concentration were prepared and
analyzed thrice a day (morning, afternoon, and evening). The result was
indicated by % RSD.
LOD and LOQ:
The LOD is the detection limit of an individual analytical procedure is
the lowest amount of analyte in a sample, which can
be detected, but not necessarily quantitated as an
exact value. The LOQ is the concentration that can be quantitated
reliably with a specified level of accuracy and precision. The LOD & LOQ
were calculated using the formula involving the standard deviation of response
and the slope of the calibration curve.
LOD= 3.3 x σ/s
LOQ= 10 x σ/s
Where
σ - Standard deviation of the response,
s – Slope ratio curve
RESULTS
AND DISCUSSION:
Figure: 2. Colorimetry
absorption spectra
Table: 1. Linearity
data
|
S.No. |
Concentration (µg/mL) |
Absorbance |
|
1 |
5 |
0.1064 |
|
2 |
10 |
0.2087 |
|
3 |
15 |
0.3435 |
|
4 |
20 |
0.4609 |
|
5 |
25 |
0.5787 |
|
6 |
30 |
0.6945 |
|
7 |
35 |
0.8212 |
Figure: 3.
Calibration graph
Figure: 4. Stability
indicating graph
Table: 2. Optical
characteristics
|
Parameter |
Result |
|
Absorption maxima λmax (nm) |
763nm |
|
Beer’s law limits (µg/mL) |
5-35µg/mL |
|
Standard regression equation
Y=mx+c |
Y=0.0236x-0.0113 |
|
Correlation coefficient |
0.999 |
|
Sandell’s sensitivity (mg/cm2/0.001 absorbance
unit) |
0.04733 |
|
Molar absorptivity
(lit.mol-1cm-1) |
0.4771×105 |
|
LOD |
0.769 µg/mL |
|
LOQ |
2.33µg/mL |
|
Assay (n=6) |
99.23 |
Table: 3. Recovery
studies of Tapentadol hydrochloride
|
Name |
Declared value (µg/mL) |
Spiked value (µg/mL) (n=3) |
Amount found (µg/mL) (n=3) |
% Recovery ± S.D |
|
Tapentadol HCl |
10 10 10 |
8 10 12 |
17.89 19.79 22.26 |
99.90±0.37 99.58±0.46 100.51±0.55 |
Table: 4. Precision results
showing repeatability of Tapentadol Hydrochloride
|
Amount of
drug taken (µg/mL) |
Average
absorbance** ± S.D |
%R.S.D |
|
10 |
0.2081±0.00186 |
0.89 |
**Average of six readings
Table: 5. Results for Intermediate Precision of tapentadol
HCl for inter and intraday study
|
Amount of drug (µg/mL) |
Average absorbance in intraday studies** (µg/mL) |
Average absorbance in inter day studies** (µg/mL) |
Precision (intraday) |
Precision (inter day) |
||||||
|
10 |
Sess-I |
Sess-II |
Sess-III |
1st
day |
2nd
day |
3rdday |
S.D |
%
R.S.D |
S.
D |
%R.S.D |
|
0.2094 |
0.211 |
0.2098 |
0.212 |
0.210 |
0.214 |
0.00068 |
0.32 |
0.0016 |
0.77 |
|
**Average of six readings
The proposed method was simple, rapid and precise. It does not
suffer from any interference due to common excipients
of tablet. In the presence of alkaline medium, tapentadol
hydrochloride reacts instantaneously with the FCR resulting in blue colored
product. Formation of blue color product is due to a common reaction mechanism
i.e., the oxidation of the drug and the reduction of FCR. Folin ciocalteu reagent is a mixture of acids and involves the
following chemical species:
3H2O.P2O5 13WO3.5MoO3.10H2O
and 3H2O.P2O5.14WO3 .4MoO3 .10 H2O
The blue color
formation by FCR
with tapentadol
HCl seems analogous
to Folin phenol
protein reaction. Tapentadol HCl probably
reduce tungstate and/or
molybdate
in FCR producing reduced
species which have
a characteristic intense blue
color with λ max at 763 nm. The
stability of colored
complex was checked
with respect to time and that color
was stable for
1.0 hour (Figure: 4), so it is
recommended that the readings
should be taken
within the specified
time range. The proposed method obeyed beer’s law in the
concentration range of 5-35µg/mL with FC reagent
respectively. Linearity data for the proposed method is incorporated in Table: 1. the optical characteristics
and the data concerning the proposed method were represented in Table: 2. The sensitivity was estimated
in terms of limit of quantification (LOQ), the smallest amounts detected were
estimated in terms of limit of detection (LOD) The results also shown in Table:2. The recovery studies were
carried out for the developed method by the addition of standard drug solution
of Tapentadol to pre-analyzed solution. The recovery
studies were satisfactory and the percentage of drug recovered (Table: 3) was in the range of
99.90-101.51%, which showed that there was no interference from excipients. The precision of the method expressed as % RSD
of intraday and interday validation is given in Table: 4 and 5. The method was validated and found to be simple,
sensitive, accurate, and precise. The study has acceptable correlation
coefficient and % RSD. It was
successfully applied for determination of drugs in their newly formulated
pharmaceutical formulations.
CONCLUSION:
This is for the first time that
spectrophotometric methods are being reported for the assay of Tapentadol Hydrochloride in pure form and also in its
formulations. Unlike GC and HPLC Techniques, Spectrophotometry is simple and
inexpensive. The Spectrophotometric methods require simple reagents, which an
ordinary analytical laboratory can afford and the Procedures do not involve any
critical reactions. Moreover the proposed method is found to be simple,
sensitive, accurate, and with good precision. The amounts obtained by the
proposed methods are between 89% and 111% and the acceptance criteria
is between 85% and 115%. The recovered percentage values ranges between
99.90 and 101.51. Thus, this approach could be considered for the analysis of
this drug in the quality control laboratories.
ACKNOWLEDGEMENTS:
The authors are thankful to the Lupin
Pharmaceuticals for providing the gift sample of tapentadol. The authors are also thankful to the management of
Hindu college of Pharmacy for providing necessary facilities to carry out this
project.
REFERENCES:
1.
Neil
M.J.O. (Ed. By) The merck
Index – An Encyclopedia of chemicals, Drugs an Biological.Merck and Co. Inc., 13th Edn. 3626; P, 643.
2.
Tzschentke TM et al. (1R,2R)-3-(3-Dimethylamino-ethyl-
2- methyl-propyl)-phenol Hydrochloride (Tapentadol HCl) , a Novel μ-Opioid
Receptor Agonist/Norepinephrine Reuptake Inhibitor
with Broad-Spectrum Analgesic Properties. Journal of Pharmacology and Experimental
Therapeutics. 323(1); 2007: 265-76.
3.
Daniels, S et al.
A Randomized, Double-Blind, Phase III Study Comparing Multiple Doses of Tapentadol and Placebo for Postoperative (bunionectomy) Pain. Current Medical Research and Opinion. 25
(3); 2009: 765-776.
4.
James A Bourland, Ayodele A Collins and
Scot A Chester. Determination of Tapentadol (NUCYNTA)
and N- desmethyl Tapentadol
in authentic urine by UPLC and Tandem Mass Spectroscopy. Journal of analytical toxicology. 34; U.S.A, 2010.
5.
Snell F.D, Snell
F.T, and Van Nostrand.D. Colorimetric Methods of Analysis,
Toronto, N. Y., London. III (III); 1953: 104.
6.
ICH. Note for
guidance on validation of analytical methods: Definitions and Terminology Q2A;
1994.
7.
ICH. Note for
guideline Validation on Analytical Procedures: Methodology Q2B (ICH, Geneva,
Switzerland, 1996).
8.
ICH, Q2 (R1)
Validation of Analytical Procedures: text and methodology; 2005.
Received on 10.10.2012
Modified on 17.10.2012
Accepted on 25.10.2012
© A&V Publication all right reserved
Research Journal of Pharmaceutical Dosage
Forms and Technology. 4(6): November–December, 2012, 328-331